PROGRAMMING AN INTERMITTENT THERAPY USING MANUAL PROGRAM MODE
NOTE: When operating the pump on battery power, the LCD screen dims after a period of time to
conserve battery power. Any keypress will illuminate the screen.
NOTE: Use the UP/DOWN ARROW keys to move highlighter. Use the numeric keys for data entry.
1. Turn the pump ON and wait for the self-test to complete.
2. Highlight PROGRAM from the SELECT menu and press YES.
NOTE: The following screen (“NEW PROGRAM”) will not appear if the previous RX has been cleared from
the pump. If this screen does not appear, skip to step #4.
3. Highlight NEW PROGRAM and press YES. Enter access code, if prompted.
4. Highlight INTERMITTENT from the therapy selection menu and press YES. If any on-screen alerts
appear, press the YES key.
5. With Unidentied highlighted, press YES (this indicates Manual Programming Mode).
6. Select units: Press YES if units desired is indicated (choices are mg, mcg or mL). To change units
press NO until desired unit is reached, then press YES to conrm. If mg or mcg is selected, enter the
concentration and press YES.
7. Press YES to leave the Delayed Start O, or press NO to set a delay and refer to the Delayed Start
section.
8. At Next? Press YES to continue.
9. Enter Bag Volume, press YES.
NOTE: To force an auto-calculation, enter any 2 of the 3 entries: AMT/DOSE, DOSE RATE or DOSE
DURATION, the unknown eld will auto-calculate. Bypass the unknown eld by using the DOWN
ARROW key. Enter the known elds pressing YES to conrm each. The unknown eld will auto-
calculate. Press YES to conrm this eld.
10. Enter AMT/DOSE (if known), press YES.
11. Enter DOSE RATE (if known), press YES.
12. Enter DOSE DURA (if known), press YES.
13. Continue pressing YES if needed until the DOSE Freq eld is highlighted.
14. Enter DOSE Freq, press YES.
NOTE: If a KVO rate of 0.1 to 10 mL is entered, the pump will give audible alerts every 20 minutes at the
end of the infusion. Whereas if zero KVO is entered the pump will alarm every 3 seconds at the
end of the infusion.
15. Enter KVO rate (if any), press YES.
16. DOSE/BAG is auto-calculated. Press YES to conrm or NO to change, input desired doses then press
YES.
17. The next 2 elds are auto-calculated: REQ VOL and TOT Time. Press YES to conrm each eld.
18. Press YES at DONE?
INTERMITTENT THERAPY

Other manuals for Moog curlin 6000 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Moog curlin 6000 series and is the answer not in the manual?

Moog curlin 6000 series Specifications

General IconGeneral
BrandMoog
Modelcurlin 6000 series
CategoryMedical Equipment
LanguageEnglish

Related product manuals