PROGRAMMING A TPN THERAPY USING PROTOCOL LIBRARY MODE
NOTE: When operating the pump on battery power, the LCD screen dims after a period of time to
conserve battery power. Any keypress will illuminate the screen.
NOTE: Use the UP/DOWN ARROW keys to move highlighter. Use the numeric keys for data entry.
1. Turn the pump ON and wait for the self-test to complete.
2. Highlight LIBRARY from the SELECT menu and press YES.
3. Highlight NEW PROGRAM and press YES. Enter access code, if prompted.
NOTE: While YES is highlighted in the Page Down eld, press YES to move to the next page while in the
CAT (category) and DRG (drug) menus.
4. Highlight the desired CAT, press YES.
5. Highlight the desired DRG, press YES.
6. Press YES to ACCEPT?
7. If any on-screen alerts appear, press the YES key.
8. Note any on-screen ADVISORY and press YES to continue.
9. Press YES to leave the Delayed Start O, or press NO to set a delay and refer to the Delayed Start
section.
10. At Next? press YES to continue.
11. Accept pre-programmed values for BAG VOL, Vol TBI, INF RATE, UP Ramp, DN Ramp, TOT TIME and
KVO if they are correct, by pressing YES to each eld.
12. If changes are needed to the pre-programmed values, highlight the desired eld and press NO.
Enter the correct value and press YES.
13. Continue as above until programming is completed.
14. Press YES at DONE?
SETTING A DELAYED START
1. At the Delay eld, press NO to OFF, then press YES to ON.
2. Enter Time infusion should begin (HH:MM), press YES.
3. Enter Date infusion should begin (MM/DD/YY), press YES.
4. Press YES to ACCEPT.
TPN THERAPY
“RUN TO START” SCREEN
1. Press the NO key to review the program and/or change the RX (see CHANGING THE RX section for
instructions).
2. Press the OPTIONS Key to review or change desired pump settings (see OPTIONS section for
instructions).
3. Press the PRIME Key to prime the set (see Priming section under GETTING STARTED for instructions).
4. Press RUN to start the infusion.

Other manuals for Moog curlin 6000 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Moog curlin 6000 series and is the answer not in the manual?

Moog curlin 6000 series Specifications

General IconGeneral
BrandMoog
Modelcurlin 6000 series
CategoryMedical Equipment
LanguageEnglish

Related product manuals