Note: The Moog Curlin Infusion pump requires conrmation to each entry to move to the next eld or
screen.
HELP/OPTIONS: Used to request a help screen when the pump is paused or at the “Infusion Complete
screen. Help screens correspond to words in the highlight bar. Use the UP/DOWN ARROW keys to
highlight the selection needing HELP. It is also used to access the OPTIONS menu when the pump is
running. When the OPTIONS menu is displayed, help is again available with this key. When in a help
screen, use the arrow keys to scroll through the help message. Press the HELP key again to exit Help.
Refer to the OPTIONS section of the individual therapy modes within this booklet for assistance with
OPTIONS.
./SILENCE: Used as a decimal point when programming numeric data. Used also to temporarily silence
the alarm (for one-minute intervals).
PRIME/BOLUS: Used to prime uid through the tubing to clear it of air or to administer a patient
demand bolus dose of medication in PCA mode.
0 - 9 Numeric keys: Used for data entry during the programming screens and the display of
Information-On-Demand (IOD) functions when the pump is in the RUN mode.
INFORMATION ON DEMAND IOD KEYS:
To view any of the Information-On-Demand (IOD) screens on the pump display, press the related key
while the pump is infusing (running a therapy). The information requested will be displayed for several
seconds and then the display will return to the previous infusion screen.
The rst row of numbers (1 through 4) function as IOD in all 5 therapy modes.
The second and third rows of numbers (5 through 0) function as IOD in the PCA Mode ONLY.
Below is the designation and functionality of the ten numeric keys of the IOD keyboard:
IOD FUNCTION FOR ALL THERAPIES:
1. Date and Time (DateTime): The Date and Time screen displays the current date and time according to
the pump.
2. Power Check (PowrChck): The Power Check screen displays the active power source of the pump,
whether internal (batteries) or external. If the active source is the C-cell batteries, a bar graph will
indicate the amount of power remaining in the batteries.
3. Option Info (OptnInfo): This screen provides information on 4 selected options. The information
displayed here includes the status of the Lock Level, Down Occlusion alarm sensitivity, Air-in-Line
alarm sensitivity, and the Audio Level. Changes to this screen are not allowed.
4. Hold Screen (HoldScrn): Pressing the Hold Scrn (#4) key will hold the current Run screen information
for an additional 8 seconds and then the pump will return to the normal running screen. The hold
screen functions for the Infusion Running Screen only.

Other manuals for Moog curlin 6000 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Moog curlin 6000 series and is the answer not in the manual?

Moog curlin 6000 series Specifications

General IconGeneral
BrandMoog
Modelcurlin 6000 series
CategoryMedical Equipment
LanguageEnglish

Related product manuals