PROGRAMMING A VARIABLE THERAPY USING PROTOCOL LIBRARY MODE
NOTE: When operating the pump on battery power, the LCD screen dims after a period of time to conserve
battery power. Any keypress will illuminate the screen.
NOTE: Use the UP/DOWN ARROW keys to move highlighter. Use the numeric keys for data entry.
1. Turn the pump ON and wait for the self-test to complete.
2. Highlight LIBRARY from the SELECT menu and press YES.
3. Highlight NEW PROGRAM and press YES. Enter access code, if prompted.
NOTE: While YES is highlighted in the Page Down eld, press YES to move to the next page while in the CAT
(category) and DRG (drug) menus.
4. Highlight the desired CAT, press YES.
5. Highlight the desired DRG, press YES.
6. Press YES to ACCEPT?
7. If any on-screen alerts appear, press the YES key.
8. Note any on-screen ADVISORY and press YES to continue.
9. Press YES to conrm the dosing UNITS.
10. If a concentration is present, ensure the concentration matches the medication container. If correct,
press YES to conrm.
11. Press YES to leave the Delayed Start O, or press NO to set a delay and refer to the Delayed Start section.
12. At Next? press YES to continue.
13. Accept pre-programmed values for BAG VOL, # DOSES and KVO if they are correct, by pressing YES to
each eld.
14. If changes are needed to the pre-programmed values, highlight the desired eld and press NO. Enter
the correct value and press YES.
15. Continue as above until NEXT? Is highlighted.
16. At NEXT? press YES to continue.
17. Accept pre-programmed values for Amt TBI, RATE and TIME for each dose if they are correct, by pressing
YES to each eld.
18. If changes are needed to the pre-programmed values, highlight the desired eld and press NO. Enter
the correct value and press YES. At the NEXT? eld in each dose, press YES to continue.
19. Press YES to each value in the summary screen: TOT DOSES, TOT TIME, REQ VOL.
20. Press YES at DONE?
SETTING A DELAYED START
1. At the Delay eld, press NO to OFF, then press YES to ON.
2. Enter Time infusion should begin (HH:MM), press YES.
3. Enter Date infusion should begin (MM/DD/YY), press YES.
4. Press YES to ACCEPT.
VARIABLE THERAPY