CAUTION:
Previous RX must be cleared from the pump prior to new patient use. Refer to the “CURLIN 6000 TRAINING &
QUICK REFERENCE GUIDE FOR BETWEEN PATIENT USE” (Contact Customer Service for information).
INSTALLING/CHANGING THE BATTERIES
1. PAUSE the pump.
2. Turn the pump OFF.
3. Open the battery door by sliding the gray “slider” to the right to the “UNLOCK” position and slide the
door down to remove.
4. Remove old batteries from the top rst. Insert new batteries bottom or at side against the coils rst,
then push the top side into place.
5. Replace door and return slider to “LOCK” position and turn pump ON.
6. Press YES to PROGRAM (or LIBRARY if using the protocol LIBRARY).
7. Press YES to RESUME (or YES to Repeat RX if current container has not begun or if changing the
container).
8. Press RUN to start.
INSTALLING THE ADMINISTRATION SET
1. Open pump door.
2. Remove the old tubing.
3. Remove the new administration set from the package.
4. Twist and remove breakaway tab from YELLOW Flow-Stop.
5. Locate the BLUE and YELLOW arrows inside the pump as a reference point.
6. Insert the tubing using the reference points.
7. Ensure the tubing passes through the notch in front of the door hinge.
8. Center tubing over pumping ngers.
9. Close the door securely. Be sure top of door is ush with top of the pump.