OPTIONS FOR TPN THERAPY
NOTE: Press YES at each eld to advance to the next eld. Pressing YES to ACCEPT OPT at the end of
the menu is mandatory to conrm all changes made in the OPTIONS menu.
Lock: This is the security level setting. Press YES to accept or NO to change Lock Setting. Enter access
code if prompted. Press YES to accept new setting.
HRLY TOT: This is the hourly infusion totals. Press YES to leave the screen o or NO to change to ON,
then press YES to ON to view. Use the UP/DOWN ARROW keys to view the previous 24 hour totals.
Press NO to exit the screen.
CLR HRLY: Press YES to leave the screen o or NO to change to ON, then press YES to ON to view. At
the ACCEPT? YES/NO eld, press YES to clear or NO to retain the Hourly Totals and exit the screen.
DN Ramp: Press YES to leave the screen o. If an immediate down ramp is needed, press NO to OFF
and YES to ON to access early down ramp feature. Press YES to maintain current DN Ramp time or NO
to change the amount of down ramp time, then press YES to accept.
DN Occlu: This is the downstream occlusion sensitivity setting (default setting is LOW). Press YES to
accept or NO to change then press YES to accept.
AIR SENS: This is the air-in-line sensitivity setting (default setting is 0.5 mL). Press YES to accept or NO
to change then press YES to accept.
AUDIO: This is the audible volume control: Press 1-9 to hear level, press YES to accept the desired
volume.
POWER CK: This is the power source with power graph. Press YES to leave the screen o or NO
to change to ON, then press YES to ON to view the power graph. Screen automatically returns to
OPTIONS menu after several seconds.
ACCEPT OPT: Press YES to accept any changes made. Pressing NO will cancel any setting changes.
TPN THERAPY
CHANGING THE RX (BAG VOL, Vol TBI, INF RATE, UP Ramp, DN Ramp, TOT TIME, KVO)
1. Press the PAUSE key.
2. Press YES to RESUME.
3. Press NO to review.
4. Press YES to each eld until the value to change is reached.
5. At the desired eld(s), enter the new program parameters (enter access code if prompted). If an
access code was needed, enter the new parameters again. Press YES to conrm.
6. Press YES to all other highlighted areas if no other changes are to be made.
7. At DONE? press YES.
8. Press RUN to start the infusion.