IOD FUNCTION FOR PCA THERAPY ONLY:
5. Hourly Total (HrlyTot): The information displayed here includes the hourly Total Infused (amount), the
number of Bolus Given (BG) and the number of Bolus Attempted (BA) for up to 24 previous hours.
The user can scroll up or down the screen to view the hourly infusion data by using the UP/DN
ARROW keys.
6. Therapy Information (ThrpyInfo): The information displayed in this screen is from the pre-prescription
menu and includes Units (dosing units), Concentration (of the medication), Administration Route and
Load Dose.
7. RX Information (RXInfo): This screen provides prescription menu information and includes Basal Rate,
Patient Bolus, Bolus Interval, and Number of Bolus per Hour.
8. Percent Bolus (%Bols): This screen provides Bolus Volume as the percentage of Total Volume. TBOL
is the total volume infused through bolus, TVOL is the total volume infused, and % BOLS is the
percentage of TBOL over TVOL.
9. Clear Shift Total (CLRShft): Two Shift Total screens are displayed for 11 seconds when this key is
pressed. This key allows the user to view only or CLEAR the Shift Totals. TO CLEAR SHIFT TOTALS: Press
the CLR Shft (#9) key. Record needed information (amount inf, volume inf, clin doses, boluses given
of boluses attempted). Then press the YES/ENTER key while “CLEAR” is displayed on the screen. If
“CLEAR” is not displayed on the screen, the key has timed out and returned to the Infusion Running
Screen. Press CLR Shft key again and press YES. Press the CLR Shft key one last time to verify that
the values are reset or cleared and the Elapsed Time reset to 1 minute.
0. Check Maximum Infusion (ChckMax): The information displayed here is the maximum (rate or
amount) that can be infused hypothetically, given the prescription parameters. This “Maximum
Infusion per Hour” serves as a “sanity check” after the pump has begun infusing.