PROGRAMMING A CONTINUOUS THERAPY USING MANUAL PROGRAM MODE
NOTE: When operating the pump on battery power, the LCD screen dims after a period of time to
conserve battery power. Any keypress will illuminate the screen.
NOTE: Use the
UP/DOWN ARROW
keys to move highlighter. Use the numeric keys for data entry.
1. Turn the pump ON and wait for the self-test to complete.
2. Highlight PROGRAM from the SELECT menu and press
YES
.
NOTE: The following screen (“NEW PROGRAM”) will not appear if the previous RX has been cleared from
the pump. If this screen does not appear, skip to step #4.
3. Highlight NEW PROGRAM and press
YES
. Enter access code, if prompted.
4. Highlight CONTINUOUS from the therapy selection menu and press
YES
. If any on-screen alerts appear,
press the
YES
key.
5. With Unidentied highlighted, press
YES
(this indicates Manual Programming Mode).
6. Select units: Press
YES
if units desired is indicated (choices are mg, mcg or mL). To change units
press
NO
until desired unit is reached, then press
YES
to conrm. If mg or mcg is selected, enter the
concentration and press
YES
.
7. Press
YES
to leave the Delayed Start O, or press
NO
to set a delay and refer to the Delayed Start
section.
8. Press
YES
to leave MedLMTS O, or press
NO
to set MedLMTS. Refer to the User Manual for MedLMTS
instructions.
9. At Next? Press
YES
to continue.
NOTE: Bag Volume should include overll volume if pump priming and/or a KVO is programmed.
10. Enter Bag Volume, press
YES
.
NOTE: To force an auto-calculation, enter any 2 of the 3 entries: Amt TBI, RATE or TIME, the unknown eld
will auto-calculate. Bypass the unknown eld by using the
DOWN ARROW
key. Enter the known
elds pressing
YES
to conrm each. The unknown eld will auto-calculate. Press
YES
to conrm
this eld.
11. Enter Amt TBI (if known), press
YES
.
12. Enter RATE (if known), press
YES
.
13. Enter TIME (if known), press
YES
.
14. Continue pressing
YES
if needed until the KVO eld is highlighted.
NOTE: If a KVO rate of 0.1 to 10 mL is entered, the pump will give audible alerts every 20 minutes at the
end of the infusion. Whereas if zero KVO is entered the pump will alarm every 3 seconds at the end
of the infusion.
15. Enter KVO rate (if any), press
YES
.
16. Press
YES
at DONE?
CONTINUOUS THERAPY

Other manuals for Moog curlin 6000 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Moog curlin 6000 series and is the answer not in the manual?

Moog curlin 6000 series Specifications

General IconGeneral
BrandMoog
Modelcurlin 6000 series
CategoryMedical Equipment
LanguageEnglish

Related product manuals