Operation 2017-02-24
Before making settings, check that the Ready indicator is lit.
Exposures cannot be taken when you are in the setting mode.
To enter the setting mode, press and hold the kV key (approximately ve seconds) until you hear beeps.
The timer display will show a blinking “5” (default).
Sets the amount of time before entering sleep mode
(OFF to 60 minutes).
Press the up or down key to switch the value in increments
of ve minutes.
* The default is 5 (minutes).
Selects 60 kV, 70 kV, or previous setting.
Press this key to switch the kV indicators.
60 kV lit up: Starts with the 60 kV setting
70 kV lit up: Starts with the 70 kV setting
Both 60 kV and 70 kV lit: Starts with the setting applied at
the previous usage
Selects lm photo, digital photo, or previous setting.
Press this key to switch the setting.
Light OFF: Starts with lm radiography
Light ON : Starts with digital radiography
Blink: Starts with the setting applied at the previous usage
Up or Down Key
kV Indicator
Startup settings for radiography conditions and sleep mode time
(4) User Setting
Hold (approx. 5 seconds)