28
Operation 2018-10-22
Technical Specications
Model V080
Type EX-2
This unit is X-ray generator which includes X-ray source assembly and high-voltage generator.
Protection against Electric Shock Class I, Type B
Protection against Ingress of Liquids IPX 0
Operating Altitude 3
,
000 m (max.)
Pollution Degree 2
Overvoltage Category II
Disinfection Methods By wiping with Ethanol for Disinfection (Ethanol 70 vol% to 80 vol%).
Mode of Operation Non-continuous operation
Type B Applied Part Head rest (no conductive connection to patients)
Intended Use
Model V080 is a Dental X-ray unit using with an intra-oral X-ray image receptor. Model V080 is used for dental radiographic
examination and diagnosis of teeth, jaw and oral structure by exposing an X-ray image receptor to ionizing radiation.
To obtain diagnostic images, an appropriate imaging receptor shall be used with this device.
- CCD/CMOS type receptor: Detector line pair resolution 5 lp/mm or more, low contrast resolution diameter 1.0 mm or more.
- Imaging Plate type receptor: Scan resolution 5 lp/mm or more, low contrast resolution diameter 1.0 mm or more.
- Film: ISO speed of D or higher. E or F recommended.
For the European market, the receptor shall bear the CE marking.
X-ray Tube
Manufacturer / Model Toshiba or Canon Electron Tubes & Devices / D-0711SB
Focal Spot 0.7 (IEC 60336)
Target Angle 16 degrees
Target Material Tungsten
Inherent Filtration At least 1.0 mmAl
Nominal Anode Input Power 940 W (1 sec.)
Maximum Anode Heat Content 7 kJ (10 kHU)
Nominal Tube Voltage 70 kV
Circuit (Center-Grounded) Constant Potential (DC)
Maximum Filament Current 3.0 A
Filament Voltage 3.0 V – 3.7 V (At max. Filament Current 3.0 A)
Filament Frequency Limits 0 kHz – 20 kHz
Specications
[Specications may be changed without notice due to improvements.]