29
Operation 2018-10-22
Generator/X-ray Head Assembly
Model V080
Operating Tube Potential 60/70 kV (accuracy : setting values ±10%)
Operating Tube Current 7 mA xed (accuracy : ±10%)
Reproducibility of Air Kerma Coefcient of variation max. 0.05
Maximum Output Power 490 W (70 kV, 7 mA, 0.1 sec.)
Total Filtration Min. 2.3 mmAl, 70 kV/HVL 3.1 mm.
(Fixed added lter; minimum 1.3 mmAl, X-ray tube ltration; 1.0 mmAl)
Beam Quality HVL minimum 1.5 mmAl at 70 kV
Protective Shielding 0.3 mm Pb or equivalent
Duty Cycle 1:29 (70 kV, 7 mA, for example 2 sec. exposure per 58 seconds cool-down period*)
* The maximum exposure time is 2 sec.
* The minimum interval is limited to 15 seconds to protect the hardware including
the X-ray tube.
Filament Preheated
Rectication Direct Current
Cooling Oil cooling
Maximum Heat Unit of X-ray Head 62.8 kJ (1 HU=1.35 Joule, 1 J = 1 Ws)
Maximum continues input 4.9 W
Leakage Radiation Max. 0.25 mGy/h at 1 m
Source Skin Distance (SSD) 200 mm (Standard cone)
300 mm (Long cone) (Option)
X-ray Field Size Max. 60 mm diam. at the end plane of the cone
Reference Axis Runs axially with the cone
Weight of X-ray Head Approximately 3.5 kg
Outer Dimension of X-ray Head Approximately 280 × 220 × 92 mm (Standard cone)
Approximately 380 × 220 × 92 mm (Long cone)
Max. Temperature of the Cone Approximately +48°C at the ambient temperature of +40°C.
Irradiation Time Range
0.01 to 2.0 sec. (series R’10 IEC60601-1-3: 2008 Annex B)
Accuracy Setting value ±5% or ±20 ms
Minimum mAs 0.07 mAs
Specications