Operation 2018-10-22
Taking an exposure
Ask the patient not to move during the exposures.
The operator must stay outside the X-ray room or protect him/herself from exposure if staying in the room.
Check that the Ready indicator lights up in green.
Press the Emission button or key to take an exposure.
During exposure, the exposure buzzer is heard and the exposure indicator lights up.
Hold the Emission button or key until the exposure buzzer stops.
Emission Key
Exposure Indicator
• Everyone except the patient must stay outside the X-ray room during X-ray emission. If the unit is not enclosed by an X-ray booth or other protective barrier,
everyone except the patient must stay outside from a distance not less than 2 m from tube head during X-ray emission.
• When it is required to stay in the X-ray room during the radiography, the operator must wear X-ray protective clothing.
• Always monitor the patient and the unit during X-ray radiography. Should an emergency occur, release your nger from the Emission button or key to stop the
unit.
• Hold the Emission button or key until the radiography is completed. If the Emission button or key is released before the radiography is completed, the radiogra-
phy process is interrupted.
• In the event that the unit does not respond to key operation, turn off the unit and have the patient leave the room. Power the unit on again after one minute or
more has elapsed after the unit was turned off.
If you release the Emission button or key before exposure is completed, the radiography process is interrupted and the error message “E00” is
displayed on the timer display.
To clear the message, press any key (except the Emission key) on the control panel.
Emission Button