37
Operation 2018-10-22
Electromagnetic Disturbances (EMD)
Guidance and Manufacturers Declaration – Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specied below.
The customer or the user of this device should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment – Guidance
Conducted disturbance
CISPR 11
Group 1
Class B
This device uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
Radiated disturbance
CISPR 11
Group 1
Class B
This device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Harmonic current
IEC 61000-3-2
Class A
Voltage uctuations and
icker
IEC 61000-3-3
Clause 5
The Veraview iX (hereafter “this device”) conforms to IEC 60601-1-2:2014 Ed. 4.0, the relevant international standard for
electromagnetic disturbances (EMD).
The following is the “Guidance and Manufacturer’s Declaration” which is required by IEC 60601-1-2:2014 Ed. 4.0, the relevant
international standard for electromagnetic disturbances.
This is a Group 1, Class B product according to EN 55011 (CISPR 11).
This means that this device does not generate and/or use internationally radio-frequency energy, in the form of electromagnetic
radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose and that it is suitable
for use in domestic establishments and in establishments directly connected to a low voltage power supply network which
supplies buildings use for domestic purposes.
The use environment of this device is the Professional healthcare facility environment.
This device needs special precautions regarding EMD and needs to be installed and put into service according to the EMD information provided in the ACCOM-
PANYING DOCUMENTS.
Use of parts other than those accompanied or specied by J. MORITA MFG. CORP. could result in increased electromagnetic emissions or decreased electro-
magnetic immunity of this device and result in improper operation.
Do not use this device as adjacent or stacked as possible with other. When adjoining or stacking is necessary, use it after observing whether this equipment
and other equipment work properly.
Portable and mobile RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30
cm to any part of the V080, including cables specied by the manufacturer.