23
Operation 2018-10-22
Disinfection, Replacement / Optional Parts and Storage
(1) Disinfection
After each use (patient), disinfect the cone, head rest, chair seat and seat height adjust lever by wiping them with Ethanol
for Disinfection (Ethanol 70 vol% to 80 vol%).
If it is not possible to obtain Ethanol for Disinfection (Ethanol 70 vol% to 80 vol%), use one of the disinfectants listed
below; do not use any other type of disinfectant.
DÜRR DENTAL’s FD 322 quick disinfectant
DÜRR DENTAL’s FD 333 quick disinfectant
DÜRR DENTAL’s FD 360 imitation leather cleaning and care
DÜRR DENTAL’s FD 366 sensitive Rapid disinfection
Wipe the operation panel with Ethanol for Disinfection (Ethanol 70 vol% to 80 vol%).
Do not fail to turn the power switch off. This will avoid the risk of electric shocks, burns, or accidentally pressing a switch.
Wipe with ethanol
Use only Ethanol for Disinfection (Ethanol 70 vol% to 80 vol%) or a neutral detergent to clean outer surfaces. Alkali or acidic solutions, cresol liquid
soap, and other chemicals can cause discoloration and surface damage.
Dampen a soft cloth with Ethanol for Disinfection (Ethanol 70 vol% to 80 vol%) or a neutral detergent, and wring it out thoroughly. Make sure no
liquid seeps inside; this could cause mechanical or other malfunctions.
Use Ethanol for Disinfection (Ethanol 70 vol% to 80 vol%) to immediately wipe off any water, detergent or other chemicals that get on the outer sur-
faces.
Do not directly spray Ethanol for Disinfection (Ethanol 70 vol% to 80 vol%), neutral detergent or water on the unit. Make sure no liquid seeps inside;
this could cause mechanical or other malfunctions.
Do not use ozone water to clean the unit. Ozone water could damage to the unit.
Do not disinfect the clinic with ozone gas or ultraviolet light. This could damage plastic and rubber components.
When cleaning, never pull on any cables or cords.
The X-ray image receptor must be cleaned or disinfected for each patient in order to avoid cross infection between patients. For cleaning or disinfec-
tion of the receptor, follow the instructions of the receptor’s manufacturer.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Morita Veraview iX and is the answer not in the manual?

Morita Veraview iX Specifications

General IconGeneral
BrandMorita
ModelVeraview iX
CategoryDental equipment
LanguageEnglish

Related product manuals