MSA MAINTENANCE
ALTAIR 4 - Operating Manual 43
If any sensor is not to be installed, ensure that a sensor plug is
installed properly in its place.
(5) If replacing sensor filters at this time:
Carefully peel off old filters taking care not to damage the inside of
the case.
On new filters peel off backing exposing adhesive. Note proper
orientation of each filter and apply to inside of front case adhesive
side against case.
Press filter into place taking care not to damage filter surface.
(6) Reinstall the sensor gasket in the case front.
Ensure sensor gasket is properly installed.
(7) Re-install the screws.
(8) If this sensor channel was deactivated, go into the sensor setup and
turn ON the sensor.
(9) Calibrate the instrument after the sensors have stabilized.
Allow sensors to stabilize at least 30 minutes before calibration.
WARNING
Calibration is required after a sensor is installed; otherwise, the
instrument will not perform as expected and persons relying on this
product for their safety could sustain serious personal injury or death.
4.3 Cleaning the Instrument
Clean the exterior of the instrument regularly using only a damp cloth. Do
not use cleaning agents.
4.4 Storage
When not in use, store the instrument in a safe, dry place between 65°F
and 86°F (18 °C and 30 °C). After storage, always recheck instrument
calibration before use.
4.5 Shipment
Pack the instrument in its original shipping container with suitable padding.
If the original container is unavailable, an equivalent container may be
substituted.