MSA OPERATION
ALTAIR 4 - Operating Manual 29
3.4 Instrument Setup
This section describes the configuration options that are available through the
Options Setup menus. These menus can be accessed only when the
instrument is turned ON while pressing and holding the
[
]
button
(see Sections 3.2 and 7.3).
This mode can only be activated at instrument turn-ON.
The operation is as follows:
(1) Press and hold the
[
]
button while turning the instrument ON.
• Use the
[
]
and
[
]
buttons to enter the setup password.
The default password is "672".
(2) Press
[
]
button to enter the setup menus.
• Incorrect password: instrument enters the Measure mode.
• Correct password: instrument continues/beeps three times.
The password can be changed through the MSA Link software.
In the Setup mode:
• Press the
[
]
button to store chosen value or go to the next page.
• Press
[
]
button to increase values by one or toggle
a selection ON or
OFF.
• Press and hold
[
]
button to increase values by 10.
• Press
[
]
button to decrease values by one or toggle a selection ON or
OFF.
Press and hold
[
]
button to decrease values by 10.
The following options are available by pressing the
[
]
and
[
]
buttons:
• Sensor Setup (SENSOR SETUP) - see Section 3.4.1
• Calibration Setup (CAL SETUP) - see Section 3.4.2
• Alarm Setup (ALARM SETUP) - see Section 3.4.3
• Setup Time and Date (TIME SET) - see Section 3.4.4
• EXIT