MSA INSTRUMENT SAFETY
ALTAIR 4 - Operating Manual 5
1 Instrument Safety
1.1 Correct Use
The ALTAIR 4 Multigas Detector is for use by trained and qualified per-
sonnel. It is designed to be used when performing a hazard assessment to:
Assess potential worker exposure to combustible and toxic gases
and vapors as well as low level of oxygen.
Determine the appropriate gas and vapor monitoring needed for a
workplace.
The ALTAIR 4 Multigas Detector can be equipped to detect:
Combustible gases and certain combustible vapors
Oxygen-deficient or oxygen-rich atmospheres
Specific toxic gases for which a sensor is installed.
Note: While the instrument can detect up to 25 % oxygen in ambient air, it is
approved for use only up to 21 % oxygen.
WARNING
Read and follow all instructions carefully.
Perform a bump test before each day's use and adjust if
necessary.
Perform a bump test more frequently if exposed to silicone,
silicates, lead-containing compounds, hydrogen sulfide, or high
contaminant levels.
Recheck calibration if unit is subjected to physical shock.
Use only to detect gases/vapors for which a sensor is installed.
Do not use to detect combustible dusts or mists.
Make sure adequate oxygen is present.
Do not block sensors.
Have a trained and qualified person interpret instrument
readings.
Do not recharge Li polymer battery in a combustible atmosphere.
Do not alter or modify instrument.
INCORRECT USE CAN CAUSE SERIOUS PERSONAL INJURY OR
DEATH.
NOTE:
This equipment has been tested and found to comply with the limits
for a Class A digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against
harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful

Other manuals for MSA altair 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MSA altair 4 and is the answer not in the manual?

MSA altair 4 Specifications

General IconGeneral
Operating Temperature-20°C to 50°C (-4°F to 122°F)
Ingress ProtectionIP67
TypeMulti-gas Detector
Gases DetectedLEL, O2, CO, H2S
DisplayLCD
WeightApprox. 220g (with battery)
Sensor TypeElectrochemical (for H2S, CO, and O2); Catalytic Bead (for LEL)
Alarm TypeAudible, Visual, Vibrating
Detection RangeO2: 0-30% vol, LEL: 0-100% LEL, CO: 0-999 ppm, H2S: 0-500 ppm
Humidity Range15% to 90% RH non-condensing
CertificationsUL, CSA, ATEX, IECEx