EasyManua.ls Logo

MSA Altair Pro - Page 18

MSA Altair Pro
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1. If the gas concentration reaches or exceeds the low alarm setpoint:
The instrument will:
display and flash "LO" and "ALARM" on the LCD
enter a low alarm sequence.
The low alarm can be silenced for five seconds by pressing the
TEST button; it automatically clears once the gas level falls
below the setpoint.
2. If the gas concentration reaches or exceeds the
high alarm setpoint:
The instrument will:
display and flash "HI" and "ALARM" on the LCD
enter a High alarm sequence.
The high alarm can be silenced for five seconds by pressing
the TEST button while it is still in high alarm.
The High alarm is latching and will not reset when gas
concentration falls below the HIGH setpoint.
NOTE: To reset alarm, push the TEST button after the gas
level falls below the setpoint.
Refer to the instrument during turn ON for factory-set alarm points.
If a gas concentration exceeds an alarm setpoint, the:
Audible alarm sounds
Alarm lights flash
Vibrator activates
Alarm type displays, alternately flashing the ALARM icon
and the:
LO icon (if the low alarm setpoint was exceeded)
HI icon (if the high alarm setpoint was exceeded).
3. If the STEL (Short Term Exposure Limit) reading reaches or
exceeds the STEL alarm setpoint:
The instrument will:
display and flash "LO" AND "ALARM" on the LCD
Enter a Low alarm sequence.
The STEL alarm can be silenced for five seconds by pressing
the TEST button while it is still in alarm
The STEL alarm is non-latching and will reset when STEL
reading falls below the STEL setpoint
The STEL value can be cleared. (See FIGURE 2-5 for details.)
2-10

Other manuals for MSA Altair Pro

Related product manuals