5) Time Weighted Average, TWA.
The TWA reading displays what the instrument has calculated
since power-ON.
When this page displays, press the TEST button to clear.
"CLR" displays
TWA value is automatically reset to zero when the instrument is
turned ON.
TWA value is calculated over an eight-hour exposure
The following formula is used to calculate the TWA value:
Sum of 1-minute gas readings/ 480 minutes (8 hours) = TWA value
If the instrument is ON for less than 8 hours (480 minutes), the
balance of the minute ppm values is set to zero,
6) IR mode.
When instrument displays "IR?", press button to enter IR mode.
If IR communications are not detected for three minutes or the
TEST button is pressed, the instrument will exit this mode.
See FIGURE 2-5 for additional details.
Steel Mill Version Only:
1) Test g
as mode
2) Functional check of LCD, vibrator, LEDs and horn
3) Software version displays.
4) Gas type
5) Low Alarm setpoint ("LO" "Alarm")
6) High Alarm setpoint ("HI" "Alarm")
7) STL Alarm setpoint
8) TWA Alarm setpoint
9) Peak Reading
Toxic ("HI")
The Peak/Hi and Min/Low values can be cleared.
When this page is displayed, press the TEST button to clear.
"CLR" displays.
10) Short Term Ex
posure Limit, STL
The STL reading displays that the instrument calculated since
power-ON.
When this page displays, press the TEST button to clear.
"CLR" displays.