The instrument displays an error code if it detects a problem during
startup or operation. See TABLE 5-1 for a brief description of the
error and proper corrective action.
When an inoperative component is located by using the guidelines,
consult the "Repair Procedures" later in this chapter to see if it is a
replaceable part.
Table 5-1. Troubleshooting Guidelines
PROBLEM DESCRIPTION ACTION
DISPLAY
TOGGLES
BETWEEN:
TMP/ERR Temperature out of range Return instrument to normal
temperature range and recalibrate.
If condition persists, contact MSA
during warranty period.
AD/ERR Sensor feedback error Remove battery and recalibrate.
If condition persists, contact MSA
during warranty period.
EE/ERR EEPROM inoperative Contact MSA during warranty period
PRG/ERR Program Memory
inoperative Contact MSA during warranty period
RAM/ERR RAM inoperative Contact MSA during warranty period
LED/ERR LED inoperative Verify operation of two alarm
LEDs in alarm
VIB/ERR Vibrator inoperative Verify operation of vibrator in alarm
UNK/ERR Unknown Error Contact MSA during warranty period
UND/ERR Sensor Under range Calibrate instrument
Low battery warning Remove from service as soon as
indicator and beep possible and replace battery
every 30 seconds
Low battery Alarm Instrument is no longer sensing gas.
indicator (horn sounds, Remove from service and replace
/ERR LEDs activate) battery
SNS/ERR Sensor Error Calibrate instrument. If instrument
fails calibration, replace sensor
Instrument does Low battery Replace battery
not turn ON
MEM/RST EEPROM Data Error Calibrate instrument.
Reconfigure any custom settings
(alarm setpoints, datalog, etc.)
5-3