6 Calibration
The STL06 Service-Tool or PCE06 Software are available for convenient on-site calibration (see user
manual of STL06 Service-Tool or PCE06 Software).
There is also the possibility to exchange the Sensor against a calibrated one on site. The Sensor used
can then be calibrated by the user himself or by MSR-Electronic GmbH and then reused again.
Caution:
When handling pressurised cylinders and test gases, the hazard warnings relating to com-
pressed gases and the regulations according to TRGS 220 must be observed.
Test gas can be toxic, irritant or corrosive, therefore avoid skin contact and inhalation un-
der all circumstances!
Symptoms: Dizziness, headache and nausea.
Procedure if exposed: Remove victim to fresh air, seek medical attention.