2 Sensors 2023-11 Page 28 | 40
B: Electrochemical Sensor Element for Toxic Gases and Oxygen
SENSOR ELEMENT
Gas type and measuring range
See datasheet MC2/SC2_Tox
Measuring principle
Electrochemical
Storage time
1
Ca. 6 months
Poisoning
Electrochemical sensor elements are susceptible to poisoning by organic solvents.
Gas type
Ordering No.
Measuring range
2
Accuracy
Display
resolution
Repeatability
Plastic
housing
Stain-
less
steel
housing
(SC2)
Zero-point variation
Drift in air
t
90
time
Reaction
time
t
90
time
Reaction
time
SC2- /
MC2-X-
ppm
± % sig.
ppm
< ± % sig.
≤ sec.
≤ sec.
≤ sec.
≤ sec.
± ppm
< % /
month
CO
E1110-A
050
2
0.01
3
15
5
35
5
1
2
CO
E1110-C
0150
2
0.1
5
10
5
40
5
4
0.4
CO
E1110-E
0250
2
0.1
5
10
5
40
5
4
0.4
CO
E1110-F
0300
2
0.1
5
10
5
40
5
4
0.4
CO
E1110-H
0500
2
0.1
5
10
5
40
5
4
0.4
NH
3
E1125-A
0100
5
0.1
10
120
5
200
10
5
2
NH
3
E1125-B
0300
3
0.1
10
120
5
200
10
5
2
NH
3
E1125-C
0500
3
0.1
10
120
5
200
10
5
2
NH
3
E1125-D
01000
3
1
10
120
5
200
10
10
2
NH
3
E1125-E
05000
2
1
10
40
5
120
10
100
2
NO
E1129-C
0100
2
0.1
5
20
5
40
5
3
2
NO
2
E1130-A
010
5
0.01
2
25
5
-
-
0.2
2
NO
2
E1130-B
020
5
0.01
2
25
5
-
-
0.2
2
NO
2
E1130-C
030
5
0.01
2
25
5
-
-
0.2
2
NO
2
E1130-E
0100
5
0.1
2
25
5
120
10
2
2
NO
2
E1130-F
05
5
0.001
2
25
5
-
-
0.1
2
VOC
E1160-A
010
5
0.01
5
70
5
70
10
0.5
1
VOC
E1160-B
05
5
0,001
5
70
5
70
10
0.5
1
HCN
E1183-B
050
5
0.01
5
30
5
40
10
2
2
HCN
E1183-C
0100
5
0.1
5
30
5
60
10
2
2
CH
2
O
E1185-B
010
n.d.
0.01
5
60
5
-
-
0.2
2
HCl
E1186-D
020
5
0.01
5
60
5
60
15
0.5
2
C
2
H
4
E1189-C
0200
n.d.
0.1
5
120
5
-
-
5
5
O
3
E1190-A
05
n.d.
0.001
5
60
5
-
-
0.1
2
O
3
E1190-B
010
n.d.
0.01
5
60
5
-
-
0.2
2
Cl
2
E1193-C
010
n.d.
0.01
5
40
5
-
-
0.2
2
Cl
2
E1193-D
020
n.d.
0.01
5
40
5
-
-
0.2
2
H
2
E1194-A
01000
n.d.
1
5
70
5
90
10
10
2
SO
2
E1196-B
020
3
0.01
2
20
5
-
-
0.2
2
H
2
S
E1197-A
050
3
0.01
2
25
5
70
10
1
2
H
2
S
E1197-B
0100
3
0.1
2
25
5
70
10
1
2
H
2
S
E1197-C
0200
3
0.1
2
25
5
70
10
2
2
H
2
S
E1197-D
0500
3
0.1
2
25
5
70
10
5
2
H
2
S
E1197-E
01500
3
1
5
60
5
90
10
15
2
ETO
E1199-A
010
n.d.
0.01
5
150
10
-
-
1
2
1
If stocked for a longer period, we recommend checking the zero point and recalibrating if necessary.