User Manual PolyGard
®
2 Sensors 2023-11 Page 25 | 40
9.2 Digital Sensor SC2
5 V DC from Board, e.g. SB2/MSB2, reverse polarity protected
Power consumption
• Catalytic (Pellistor)
• Electrochemical
• Semiconductor
• Infrared Premium
• Infrared
• MPS
TM
200 mA, max. (1.0 VA)
10 mA, max. (0.05 VA)
160 mA, max. (0.8 VA)
60 mA, max. (0.3 VA)
50 mA, max. (0.25 VA)
25 mA, max. (0.13 VA)
Serial interface local bus
Flammability classification
Similar to RAL 7035 (light grey)
Type P: 24 x 22 mm (0.94 x 0.87 in.)
Type L: 24 x 30 mm (0.94 x 1.18 in.)
Type S: 30 x 61 mm (1.18 x 2.40 in.)
Screw mounting,
external thread M25 x 1.5 mm
Screw mounting,
NPT ¾“
ANSI/ B1.20.1 / M30 x 1,5 mm
Standard: Ca. 150 mm (5.91 in.)
Cable extension (5, 10 or 15 m)
Cable extension (5, 10 or 15 m)
18–29 V DC, reverse-polarity protected;
18–27 V AC (only for output signal 2–10 V)
Power consumption
• Catalytic (Pellistor)
• Electrochemical
• Semiconductor
• Infrared Premium
• Infrared
• MPS
TM
75 mA, max. (1.8 VA at 24 V)
23 mA, max. (0.6 VA at 24 V)
75 mA, max. (1.8 VA at 24 V)
75 mA, max. (1.8 VA at 24 V)
40 mA, max. (1.0 VA at 24 V)
40 mA, max. (1.0 VA at 24 V)
Proportional, overload and short-circuit proof,
load ≤ 500 Ω for current signal, ≥ 50 kΩ for voltage signal
4–20 mA or 2–10 V = measuring range
3–4 mA or 1.5–2 V = underrange
> 20–21.2 mA or 10–10.6 V = overrange
2 mA or 1 V = fault
> 21.8 mA or 10.9 V = fault High
Similar to RAL 7035 (light grey)
Dimensions (Ø x H)
• Housing type P
• Housing type L (IR sensor)
24 x 22 mm (0.94 x 0.87 in.)
24 x 30 mm (0.94 x 1.18 in.)
IP65 (only if mounted in housing type A, D)
Screw-type terminal 0.25–1.3 mm
2
,3-pin, 24 to 16 AWG