User Manual PolyGard
®
2 Sensors 2023-11 Page 26 | 40
9.4 Sensor Element
A: Catalytic Sensor Element (Pellistor) for Combustible Gases
Gas type and measuring range
See datasheet DB_MC2/SC2_Ex
Pellistor (catalytic bead principle)
-30 °C to +60 °C (-22 °F to 140 °F)
Storage temperature range
1
0 °C to +20 °C (32 °F to 68 °F)
Pellistors can be poisoned by silicone-containing substances or other catalyst
poisons up to complete loss of sensitivity.
1
A deviating storage temperature can have a negative effect on sensitivity and service life.
2
If stocked for a longer period, we recommend checking the zero point and recalibrating if necessary.
3
Expected service life under normal ambient conditions.
4
Manufacturer-recommended calibration intervals for normal environmental conditions