EasyManua.ls Logo

MSR ELECTRONIC PolyGard 2 MC2 - Sensor Element; A: Catalytic Sensor Element (Pellistor) for Combustible Gases

Default Icon
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Manual PolyGard
®
2 Sensors 2023-11 Page 26 | 40
9.4 Sensor Element
A: Catalytic Sensor Element (Pellistor) for Combustible Gases
SENSOR ELEMENT
Gas type and measuring range
See datasheet DB_MC2/SC2_Ex
Measuring principle
Pellistor (catalytic bead principle)
Temperature range
-30 °C to +60 °C (-22 °F to 140 °F)
Humidity range
095 % RH not condensing
Pressure range
90110 kPa
Storage temperature range
1
0 °C to +20 °C (32 °F to 68 °F)
Storage time
2
Ca. 6 months
Service life
3
in air
> 5 years
Calibration interval
4
6 months
Poisoning
Pellistors can be poisoned by silicone-containing substances or other catalyst
poisons up to complete loss of sensitivity.
1
A deviating storage temperature can have a negative effect on sensitivity and service life.
2
If stocked for a longer period, we recommend checking the zero point and recalibrating if necessary.
3
Expected service life under normal ambient conditions.
4
Manufacturer-recommended calibration intervals for normal environmental conditions

Table of Contents