EasyManua.ls Logo

MUNBYN IMC09 - Read Sensor Values; IR Calibration; MG;MT Waveforms

Default Icon
15 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
After input the password, the interface showed in Figure 5-3 will be
displayed.
Use the cursor to select with sub-menu you want to enter, and press
MENU button to confirm, and press ESC button to quit.
13
Support: +1 650 206 2250
14
Support: +1 650 206 2250
Figure 5-2 Password Interface
Figure 5-3 Sensor Values
Figure 5-5 IR Calibration Steps
Figure 5-6 MG/MT Waveforms
5.1.1. Read Sensor Values
As shown in Figure 5-4, the sensors values are only for engineer to check
whether any sensor has a malfunction. If you meet any problem about
use, please take a photo of this page and contact us.
5.1.2. IR Calibration
5.1.3. MG/MT Waveforms
If you meet cases about banknote detection errors, please try to calibrate
the CIS. The CIS calibration steps are shown in Figure 5-5. For more detail
instruction, we recommend you to see the reference manual or watch
the video: http://u.pc.cd/qRsctalK

Related product manuals