EasyManuals Logo

NARDA EHP-50-G User Manual

NARDA EHP-50-G
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
General Information 1-7
1.6.3 Measurements
Uncertainty of EHP-50G
(Weighted Peak mode)
The total uncertainty derived from typical contributions of linearity,
anisotropy, the worst case deviation from the nominal weighting filter,
temperature and relative humidity.
Magnetic probe
Total expanded uncertainty %
Frequency from 15 Hz to 100 kHz 4.6
Frequency from 10 Hz to 300 kHz 7.1
Frequency from 3 Hz to 300 kHz 8.6
Electric probe
)
Total expanded uncertainty %
Frequency from 15 Hz to 250 kHz 8.3
Frequency from 3 Hz to 300 kHz 9.7
Note: below 3 Hz and above 300 kHz the uncertainty increases to 0.85 dB (band limiting filter not considered).

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NARDA EHP-50-G and is the answer not in the manual?

NARDA EHP-50-G Specifications

General IconGeneral
BrandNARDA
ModelEHP-50-G
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals