General Information 1-9
A good evaluation of the anisotropy of a field probe should be carried out in
a place where all of the following conditions are satisfied:
Far Field
Plane wave
Uniform field
Since the calibration systems are not infinite, these conditions can be
approximated but never reached, and so the evaluation of anisotropy for
our probes shows results worse than the reality. So, the real typical
anisotropy response is better than what specified.
EHP-50G is housed in a small cubic case. The bottom side panel includes
an optical fiber connector, extension rod screw, battery charger connector,
ON/OFF button and Status LED.
Fig. 1-3 EHP-50G axes
The sensitive elements are located approximately 9 mm below the external
surface