9-8 Accessories
9.5.4 Main
specifications
Table 1-1 lists the specifications of 8053 DISPLAY. The specifications of all
accessories are listed in the Chapter on Accessories.
The following conditions apply to all specifications:
• Temperature for use must be between -10°C and +40°C.
TABLE 9-4 Technical Specifications of 8053 DISPLAY
Frequency range
Depending on the probe
Dynamic range
>140 dB (depending on the probe)
Operating range
|
Resolution
| Depending on the probe (See Table 1-2)
Sensitivity
|
Units
V/m, kV/m, μW/cm², mW/cm², W/m², A/m, nT, μT, mT;
LCD Display
Field measured
X, Y, Z in absolute values, percent and total.
Time
Internal clock in real time
Probe
Display of the model and date of calibration
Graphic bar
The analog bar displays:
- real time value with respect to full scale;
- field versus time (in linear or logarithmic form) with automatic time
scaling;
- alarm threshold.
Measuring function
Internal memory
Up to 32.700 measurements (up to 8.100 standard memory, up to 21.600
extended memory)
Alarm
Variable threshold from 0 to 100% of full scale. Internal sound and blinking
symbol on the display when the level is greater than the alarm threshold
Functions
Minimum, Maximum and Averaging
Averaging mode
Arithmetic, quadratic (RMS), manual, rolling and spatial
Averaging time
Definable 30 sec, 1, 2, 3, 6,10,15, 30 min or manual
Data acquisition
Sampling mode (1, 10-900 sec/sample), data change, over the limit,
(Logger)
average on 6 min, manual, spectrum (with EHP-50C)
General specifications
Output
LCD display 72x72mm 128x128 pixel, RS232 (with cable or fiber optic)
Input
Fiber optic connector
Internal battery
Rechargeable at NiMH (5 x 1.2 V)
Operational time
24 hours normal mode, 48 hours (in SAVE MODE function: display off)
Recharge time
< 4 hours (15 minutes charge for 1 hour of use)
External power supply
DC, 10 - 15 V, I = about 500 mA
Interfaces
RS232 (calibration and firmware update)
Software/Firmware
Upgrade available via Internet at the Web site: http://www.narda-sts.it
Autotest
Automatic during switch-on of all functions;
Calibration
Inside the built-in E²PROM of the probe
Conformity
With Directives 89/336 and 73/23 and the amendments to them
CEI 211-6 and 211-7
Operating temperature
From -10 to +40°C
Storage temperature
From -20 to +70°C
Size (WxHxD)
108 x 240 x 50 mm
Weight
1.07 kg
Tripod support
Threaded insert ¼”