Chapter 4 Analog Input
© National Instruments 4-23 X Series User Manual
Using Non-Referenced Single-Ended (NRSE)
Connections for Ground-Referenced Signal Sources
Figure 4-11 shows how to connect ground-reference signal sources in
NRSE mode.
Figure 4-11. Single-Ended Connections for Ground-Referenced Signal Sources
(NRSE Configuration)
AI <0..31> and AI SENSE must both remain within ±11 V of AI GND.
To measure a single-ended, ground-referenced signal source, you must use
the NRSE ground-reference setting. Connect the signal to one of
AI <0..15> and connect the signal local ground reference to AI SENSE.
You also can connect the signal to one of AI <16..31> and connect the
signal local ground reference to AI SENSE 2. AI SENSE is internally
connected to the negative input of the NI-PGIA. Therefore, the ground
point of the signal connects to the negative input of the NI-PGIA.
Any potential difference between the device ground and the signal ground
appears as a common-mode signal at both the positive and negative inputs
of the NI-PGIA, and this difference is rejected by the amplifier. If the input
circuitry of a device were referenced to ground, as it is in the RSE
ground-reference setting, this difference in ground potentials would appear
as an error in the measured voltage.
MIO X Series Device Configured in NRSE Mode
Input Multiplexers
I/O Connector
AI GND
AI SENSE
or AI SENSE 2
AI <0..15>
or AI <16..31>
–
+
–
+
V
cm
V
s
Ground-
Referenced
Signal
Source
Common-
Mode
Noise
and Ground
Potential
PGIA
–
+
–
+
V
m
Measured
Voltage
Instrumentation
Amplifier
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments NI 6323 and is the answer not in the manual?

National Instruments NI 6323 Specifications

General IconGeneral
BrandNational Instruments
ModelNI 6323
CategoryI/O Systems
LanguageEnglish

Related product manuals