PXIe-4466/4467 Calibration Procedure
12 ni.com/services
4. Start the task.
5. Acquire readings with the DUT.
6. On a per channel basis, average all of the acquired samples.
7. Stop and clear the task.
8. Compare the averaged values to the appropriate limits in Table 1. The
average is the offset at the tested gain setting.