EasyManuals Logo

National Instruments PXIe-4467 User Manual

National Instruments PXIe-4467
41 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #12 background imageLoading...
Page #12 background image
PXIe-4466/4467 Calibration Procedure
12 ni.com/services
4. Start the task.
5. Acquire readings with the DUT.
6. On a per channel basis, average all of the acquired samples.
7. Stop and clear the task.
8. Compare the averaged values to the appropriate limits in Table 1. The
average is the offset at the tested gain setting.

Other manuals for National Instruments PXIe-4467

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments PXIe-4467 and is the answer not in the manual?

National Instruments PXIe-4467 Specifications

General IconGeneral
BrandNational Instruments
ModelPXIe-4467
CategoryComputer Hardware
LanguageEnglish

Related product manuals