Chapter 7 PFI
NI USB-621x User Manual 7-4 ni.com
Connecting PFI Input Signals
All PFI input connections are referenced to D GND. Figure 7-3 shows this 
reference, and how to connect an external PFI 0 source and an external 
PFI 2 source to two PFI terminals.
Figure 7-3.  PFI Input Signals Connections
PFI Filters
You can enable a programmable debouncing filter on each PFI signal. 
When the filters are enabled, your device samples the input on each rising 
edge of a filter clock. USB-621x devices use an onboard oscillator to 
generate the filter clock with a 40 MHz frequency.
Note NI-DAQmx only supports filters on counter inputs.
The following is an example of low to high transitions of the input signal. 
High to low transitions work similarly.
PFI 0 
Source 
PFI 2 
Source 
USB-621x Device
D GND 
PFI 2 
PFI 0 
I/O Connector