Error Unresolved
We are sorry this troubleshooting did not resolve your Trident Error.
Error Condition
Test A Fail
Explanation of Error
Test A Fail
The “Test A Fail” error indicates there is an error in the titration associated with reagent A.
Recommended Course of Action
Test A Fail
This most often occurs because there is no reagent in the reagent bottle, but there is still reagent level
according to the Trident tile and Apex.
To resolve this, run the reagent A replacement task in Apex Fusion or see the guide to manually replace
reagents.
After replacing the reagent, reboot the Trident or unplug the 1LINK cable to the Trident.
Error Condition
Test B Fail
Explanation of Error
Test B Fail
The “Test B Fail” error indicates there is an error in the titration associated with reagent B.
Recommended Course of Action
Test B Fail
This most often occurs because there is no reagent in the reagent bottle, but there is still reagent level
according to the Trident tile and Apex.
To resolve this, run the reagent B replacement task in Apex Fusion or see the guide to manually replace
reagents.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Neptune Systems Trident and is the answer not in the manual?

Neptune Systems Trident Specifications

General IconGeneral
Measurement ParametersAlkalinity, Calcium, Magnesium
TypeAutomated Water Tester
CompatibilityNeptune Systems Apex Controller
Power Requirements12V DC
Weight3.5 lbs (1.6 kg)