12
|
EXPMT 2000 User Manual/R5
Transcutaneous Pacemaker Test
The
Analyzer
tests various types of transcutaneous pacemakers. Most
transcutaneous pacemakers are built into cardiac resuscitation
equipment along with defibrillators
.
Defibrillators deliver high voltage
shocks
to a patient with certain
heart
conditions
.
CAUTION
To avoid possible shock hazard and damage to the
Analyz
er
,
do
not
discharge defibrillator into the unit.
Follow the manufacturer’s user manual for recommended
test
procedures and observe safety
precautions
.
Getting Started
All pacer test functions are menu driven. It is simple and easy
t
o
use by selecting the desired functions from the menu.
50 Ohms built in load is set as default. Howe
ver
,
v
arious
m
anufacturers recommend specific loads for performing the test
.
A
user selectable load box is included with the unit (as shown
below).
For tests that require ECG simulation
,
five lead low level ECG
outputs are provided.