18
EXPMT 2000 User Manual/R5
Test Procedure
1.Select “Sens” (F2) from the Pacer Measurement Menu (as
above). The sensitivity menu will be displayed.
2.Set
the
Pacer
to AAI mode, the rate to
80BPM,
and the
current
to
10mA
.
3.The default waveform is
ECG
and the default width, is 40 mS.
4.To change the waveform press F1 and to change the
width
press F2.
5.To start the test press F4 (‘Done’).
6.Press F4 when the unit displays “Sensty” ready.
7.The Display will show
Sensty:
“Wait” for a few seconds,
then the rate and sensitivity in mv will be displayed.
8.To save the test result press “Save”.