Noise Immunity Test
The immunity test determines the pacemaker’s susceptibility
t
o
50/60 Hz interference signals. The test can be performed
with
or without
ECG
wa
v
eform
.
1. From the Pacer Test Menu select “Imun” (F4). The noise
immunity test menu will be displayed.
2. Select the desired
fr
equency
,
50Hz (F1) or 60 Hz (F4).
3. To change the amplitude of the sine
wave,
use the F4 key.
4. To enable
ECG,
select F1.
5. To change the amplitude, press F4.
Press
“ESC”
to get back to the previous menu
Ventricular Pacer Output Testing
Ventricular pacer output testing is identical to the Atrial pacer
output
testing
.
1. From the main menu select “VENT” (F4) for the
V
entricular
Menu.
2. Turn
ON
the Pacemaker and set desired
Rate
and
Amp
.
3. To begin measurement press “ESC”.
4. The
analyzer
will measure the
rate,
width, and the current and
display the
result
.
5. To save the test result, press SAVE”. The results are saved
into the memory under the equipment
ID
set in the beginning
of
the
test
.
6. To display the pulse energy or volt, press “>” key. To
displa
y
the secondary
menu,
select the F1 or F3 keys respectively.
7. To save the test result press
‘SAVE’.