EXPMT 2000 User Manual/R5
|
21
Setup
Atrial Pacer Output Measurement
1.
Connect
the Atrial and Ventricular outputs of the Transvenous
Pacemaker to the respective inputs of the Analyzer as
shown
above.
2. From the main menu select “ATR” (F3) for the Atrial Menu.
3. Turn
ON
the Pacemaker and set desired
Rate
and
Amp
.
4. To begin the measurement
press,
“ESC”.
5. The
analyzer
will measure the
rate,
width, and the current and
display the
result
.
6. To save the test result press “SAVE” key. The results are
saved into the memory under the device ID set in the beginning
of the
test
.
7. To display the energy press “>” key for the secondary menu
as shown
below
.
8. To display
energy
or “Volt” press F3 or F4 respectively.
9. Press
“ESC”
to go back to the Atrial measurement menu.