EXPMT 2000 User Manual/R5
27
Accuracy: ± 1% +0.2 mS.
Refractory: 25 – 500 mS.
Accuracy: ± 2 % or ± 25ms.
Transvenous (Atrial and Ventricular)
Test Load: 500 Ohms ± 1%
Current: 1 to 40 mA.
Accuracy: ± 5% +0. 5 mA.
Pulse Rate: 20 to 200 ppm. (Rapid Pace: 805)
Pulse Width: 0.5 to 80 mS.
Accuracy: ± 1% +0.2 mS.
Refractory: 25 – 500 mS.
Accuracy: ± 2 % or ± 25ms.
Sensitivity: 1-10 mv (Atrial), 1-20 mV (Ventricular)
Accuracy: 1 mV
Immunity Test: 50 / 60 Hz.
RS 232 Output
A computer interface is available to print test results or to save
them on a
PC.
The test results can be saved directly to a PC.
Power Requirements
The unit is powered by a Li-ion rechargeable battery
(7
.
4V/
850mA). The full charge time is approximately 3-4 hours when
the
battery is fully discharged (Low Battery
Condition)
.