The Analyzer measures and displays a pacer pulse’s
amplitude
(current), peak
voltage, rate,
width, and energy.
It also conducts pacemaker demand sensitivity tests,
measuring and displaying refractory periods, and
immunity
tests, which determine the pacemaker’s susceptibility to
50/60
Hz
inter
f
er
ence
.
Pacer Pulse Measurement
1. From the Main Menu press F2 (“TRNS”).
2. The following transcutaneous menu will be displayed
a s
below
.
3. Connect the transcutaneous pacer output cables to
the
“Trans” input of the
analyzer
directly (50
Ohms
Load) or
thr
ough the load box as shown in Figure 3.
4. Select the desired load and press “ESC”.