EasyManua.ls Logo

Netech EXPMT 2000 - Pacer Pulse Measurement

Netech EXPMT 2000
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14
|
EXPMT 2000 User Manual/R5
The Analyzer measures and displays a pacer pulses
amplitude
(current), peak
voltage, rate,
width, and energy.
It also conducts pacemaker demand sensitivity tests,
measuring and displaying refractory periods, and
immunity
tests, which determine the pacemaker’s susceptibility to
50/60
Hz
inter
f
er
ence
.
Pacer Pulse Measurement
1. From the Main Menu press F2 (“TRNS”).
2. The following transcutaneous menu will be displayed
a s
below
.
3. Connect the transcutaneous pacer output cables to
the
“Trans” input of the
analyzer
directly (50
Ohms
Load) or
thr
ough the load box as shown in Figure 3.
4. Select the desired load and press “ESC”.

Related product manuals