EXPMT 2000 User Manual/R5
|
23
3. To save the test result press “Save”.
Atrial Pacer Sensitivity Test
1. From the Atrial Measurement menu select “Sens” (F2).
2. Set the pacemaker to AAI
mode
.
3. Set the rate to 80 BPM and output current to 10 mA.
4. Disable the Ventricular output (set current to 0
mA)
.
5. The waveform type and width can be changed with the
F1
and F2 keys. (The default waveform is set to
ECG
and
the
default
width to 40 mS).
6. To start the test press F4(‘Done’).
7. Press
F4
,
when the display shows: Scentsy: “Ready”.
8. The Display will show
Sensty:
“Wait” for a few seconds,
then the rate and sensitivity in mv will be displayed.
9. To save the test result press “Save”.