Model 3100 Analyzer Instruction Manual
Page 10
Manual file name: MN-A-0005, Rev. C
Manual P/N: C5-06-4900-16-0
2.1.3.1 Remote sensor module input
All connections to the Remote Sensor Module (RSM) are made by connecting the supplied interface
cable between the analyzer and the RSM. Plug the cable connectors into the mating receptacles on the
analyzer chassis and RSM.
2.1.3.2 Alarm-1 relay output
Connections from the Alarm-1 relay contacts to the user’s process control equipment are made at
terminal block TB2 on the rear of the analyzer chassis (see Figure 10). The oxygen alarm relay contacts
are voltage-free Form C relay contacts, SPDT, 5A @ 250 VAC, 5A @ 30 VDC. Connect the terminal pins
in accordance with the schematic shown on the terminal ID label on the top of the analyzer chassis.
2.1.3.3 Alarm-2 relay output
Connections from the Alarm-2 relay contacts to the user’s process control equipment are made at
terminal block TB2 on the rear of the analyzer chassis. The oxygen alarm relay contacts are voltage-free
Form C relay contacts, SPDT, 5A @ 250 VAC, 5A @ 30 VDC. Connect the terminal pins in accordance
with the schematic shown on the terminal ID label on the top of the analyzer chassis.
2.1.3.4 Fault relay output
Connections from the Fault relay contacts to the user’s process control equipment are made at terminal
block TB2 on the rear of the analyzer chassis. The fault relay contacts are voltage-free Form B relay
contacts, SPST, 5A @ 250 VAC, 5A @ 30 VDC. Connect the terminal pins in accordance with the
schematic shown on the terminal ID label on the top of the analyzer chassis.
2.1.3.5 Heater OK relay output
Connections from the Heater OK relay contacts to the user’s process control equipment are made at
terminal block TB2 on the rear of the analyzer chassis. The heater OK relay contacts are voltage-free
Form B relay contacts, SPST, 5A @ 250 VAC, 5A @ 30 VDC. Connect the terminal pins in accordance
with the schematic shown on the terminal ID label on the top of the analyzer chassis.
Fig. 9, Label on top of analyzer chassis