18 Chapter 2 System Operation
Dial Rate
- like the
Key Rate
setting, this controls the acceleration of the dial as it is
turned. Settings are
Slow
,
Medium
, and
Fast
.
2.4.9 Save/Recall Screen
Save/Recall SAV
→
→→
→
Bin=1
RCL
Figure 9 - Save/Recall Screen
The Save and Recall functions are used to store and retrieve 6000 setup
configurations for future use. For example, a specific test setup may be saved for
later use, and then another setup may be used presently. Then, when the user desires
to perform the specific test, its setup is simply recalled.
Non-volatile memory is used for saving the instrument parameters. When a save
operation is performed, all of the parameters which are currently in effect on the
6000 are stored. The user selects a “bin” number (1 - 5) for saving the parameters.
Then, when that “bin” number is recalled, the 6000 is restarted and the parameters
are reconfigured to the previously stored values.
A special “bin 0” is reserved for the reset state. Recalling bin 0 will reset all modules
in the system to factory defaults.
The save/recall bin information will be lost upon detecting any change in the module
configuration (such as installing a new module).
2.4.10 Linking Screen
# Sr Cnd Act Tg
A
→
→→
→
1
On
n5
2 PREV
B
No CLR
C
No
Figure 10 - Link Screen
The linking screen allows the conditions of the laser module and TEC to affect and
control each other.