Chapter 7 Calibration 87
f. Input the fourth actual current (as an <nrf value>) via the
TEC:ITE
<nrf value>
command.
If, at any time prior to the last
TEC:ITE,
a command other than
TEC:ITE
or
TEC:ITE?
is sent to the 6000, the 6000 will cancel the
calibration mode and then process the command(s).
Once the
TEC:ITE
value is sent, the
OPC?
query may be used to
determine when the calibration is completed. The operation complete
flag (bit 0 of the Standard Event Status Register) may be used to trigger
an interrupt. This type of interrupt is enabled by setting bit 0 of the
Service Request Enable register and using the
*OPC
command.