PSM1700 PsimetriQ user manual
4-4
4.4 Alarm function
PsimetriQ has two independent alarms that can be read
remotely or can generate an audible sound Each of the
alarms can be triggered by comparison to one or two
thresholds:
sound the alarm if the value exceeds a threshold
sound the alarm if the value is below a threshold
sound the alarm if the value is outside a window
sound the alarm if the value is inside a window
Additionally, one of the alarms can be used to generate a
sound which varies linearly between thresholds
The value to which each alarm is applied can be any of the
measurements selected for zoom. The alarm status is also
available as a logic output on the parallel port.
To program an alarm, first select the functions for the
zoom; up to four measurements can be selected for the
display, the alarm is applied to any of them; then press
ALARM to invoke the alarm menu:
select which of the zoom functions is to be used
select the type of alarm
set the upper limit (if appropriate)
set the lower limit (if appropriate)
select whether the alarm is to be latched
select whether the alarm sounder is enabled
If the alarm latch is selected then the alarm will continue
to sound even if the value returns to within the normal
boundaries. To clear the alarm, press HOME.
The linear alarm option allows tests to be carried out even
if it is not possible to see the display. Pressing ZERO in the
alarm menu sets the upper and lower threshold to 4/3 and