PSM1700 PsimetriQ user manual
14-6
The ZOOM function can be used to select up to four
parameters from the display when in real time mode. It
has no function following a sweep.
Although it is most usual to use the PsimetriQ generator
when performing LCR measurements, there may be
circumstances where this is impractical, for example
measuring the inductance of a transformer primary
winding under load. In this case, turn off the PsimetriQ
generator (OUT menu) and the frequency reference for the
analysis is measured from channel 1. Provided that the
signal is clean enough for an accurate frequency
measurement (and for DFT analysis the frequency does
need to be accurately known), then the measurements can
be made reliably.
When using an external frequency reference there can be
no sweep function.