VI Basic Operations MICROSCOPE
6 Using the Microscope Control
- 133 -
NOTE:
* If autofocus does not work correctly, try moving the sample until a sharply
contrasting part of the sample is at center of the view field. Then reexecute
autofocus.
* Supported samples are specimen slide with cover glass. (Thickness of the
slide glass + cut piece: 0.9 mm to 1.7 mm, both ends inclusive)
* Slide glass of thickness of up to 1.2 mm are supported. (Compliant with JIS
R3703-1998 or ISO 8037)
* Cover glass of thickness of up to 0.17 mm for No. 1 and 0.18 mm for
No.1-S are supported. (Compliant with JIS R3703-1996 or ISO 8255)
* Objective magnifications x1 to x100 (NA0.03 to 0.95) are supported. (Phase
contrast, liquid immersion, polarization, Hoffman, multi-immersion, and
industrial-use objectives cannot be used.)
* The supported observation method is bright-field microscopy.
* Before using the autofocus function, please be sure to read Section 6.3,
"Autofocus Precautions."
Objective <Changes objectives>
Compatible only with the Nikon 90i incorporating a motorized nosepiece.
1
In the nosepiece column , select the nosepiece number corresponding to your desired
objective. The objective is switched to the one that is attached to the nosepiece number
you've selected.
2 After switching objectives, confirm that the selected number icon is marked with a red
border and that the nosepiece number, the objective magnification, and NA value are
displayed.
NOTE:
* Before this function can be used, you must first set up objective information
from a PC using the setup software. For more information, consult the
instruction manual for your microscope.
* If you are using an intelligent nosepiece, only the objective magnification
and NA value are displayed.
* Selecting
icon shows detailed information on the objective being
used. Select the X mark on the detailed information display screen to turn
the detailed information display screen off.
* Right-clicking on the icon representing a nosepiece number shows the
magnifications and NA values of all registered objectives. You also can
select the desired objective from within this display to change objectives.
* The magnifications and NA values of objective in use are also displayed on
the "microscope status display/microscopy setup menu" (MIC Info menu).

Table of Contents

Other manuals for Nikon DS-U1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nikon DS-U1 and is the answer not in the manual?

Nikon DS-U1 Specifications

General IconGeneral
BrandNikon
ModelDS-U1
CategoryCamera Accessories
LanguageEnglish

Related product manuals