Chapter 8 Troubleshooting
2 Image Output
- 34 -
Symptom Corrective action
The contrast is low.
Set the tone appropriately using the application software.
=> To make settings, see the instructions for the application software.
Set the black level to the appropriate value using the application
software.
=> To make settings, see the instructions for the application software.
Adjust the size of the field diaphragm so that it circumscribes the
viewfield.
Adjust the size of the aperture diaphragm to approximately 70% to
80% of the numerical aperture of the objective.
Use an appropriate filter.
When using a phase contrast microscope or an interference
microscope (two luminous fluxes or multiple luminous fluxes), insert
the specified filter. With a black and white image, contrast is improved
by inserting a green interference filter (GIF). Contrast in color images
is enhanced by inserting a filter having a complementary color to that
of the specimen.
Adjust the ambient light around the microscope.
Make the room darker, or place caps on the binocular eyepiece, for
example.
Check the contrast of the specimen itself.