Chapter 8 Troubleshooting
4 Microscope Control from the PC
- 36 -
4
Microscope Control from the PC
Symptom Corrective action
Unable to control
microscope functions.
(Or incorrect control
results.)
Use only the Nikon 80i/90i Microscope. The system cannot be used
with a microscope from a different company or other Nikon
microscope models.
Turn on the power switch for the microscope.
Wait until the microscope is ready for operation. It takes about 30
seconds for the microscope to be ready.
Connect the USB cable securely between the DS-U1 to the digital
imaging head on the 80i or the 90i microscope.
Confirm that only one hub stage exists between the PC's USB port and
the DS-U1. We strongly recommend against using more than one hub.
Use the USB supplied with the microscope to connect the microscope
to the DS-U1.
Use the application software designed for control of the Nikon 80i/90i
Microscope.
Unable to display
microscope status.
(Or incorrect status
information is
displayed.)
Make sure the microscope initial setting has been performed.
Change the registration if the objective or filter has been changed.