Chapter 1 Part Names and Microscopy Procedures (Bright-Field Set)
1.3 Bright-Field Microscopy
21
1
12
Switch to the desired objective
and view the specimen.
Adjust the field diaphragm and
aperture diaphragm each time
you change objectives.
F
ield dia
p
hra
g
m:
Set slightly narrower than the margins of
the field of view.
Ap
erture dia
p
hra
g
m:
70% to 80% of the maximum numerical
aperture (N.A.) of the objective
13
Turn off power after completing
observations.
Select the
desired objective.
Press the power
switch to the ""
position.