38 / 76 ♦ BKI16ATEX0018X ♦ htk4014a0600p_03
Function Input Range Description
1.4 User data
1.4.3 Checksum
Read only This value is used for identification of the device software version. The checksum is
tested when starting. This helps to detect any problems with the microcontroller.
1.4.4 Tag number
00000 01 This parameter assigns an identification number to the device. A text consisting of
maximum 8 ASCII characters can be entered.
1.4.5 Serial number
Read only This parameter serves to identify the respective measuring device. This number
cannot be changed and sets the address for use with HART® interfaces.
1.4.6 French command
number
Read only Factory-programmed number, to be quoted in case of warranty and service claims.
1.4.6 German command
Number
Read only Factory-programmed number, to be quoted in case of warranty and service claims.
Option**
(Descriptor)
With this function an ASCII string with max. 15 characters lenght can be entered. (only
user information)
Single rod, Twin rod, Single cable, Single
cable + counterweight,
Single cable without counterweight, Twin
cable, Twin cable + counterweight, Coaxial,
Special 1, Special 2 or Special 3
1.4.9 Probe type
As per order
Information on probe type supplied with signal converter. This is a read-only
parameter.
1.5.0 Application
1.5.1. Detection delay
Minimum value: 0mm
Maximum value: Dead Zone
As per order
This function can be used to define an area directly below the flange in which
interference reflections (e.g. from the tank nozzle) are masked. This value has to be
smaller than or equal to the dead zone (Fct. 1.1.2.).