7 Mounting and Removing the Bur Guard
Applicable Attachment
Slim Handpiece 200
Super Slim Handpiece 200
_ Removing
1
Place the Wrench for Slim/Super Slim (P200-1ARA-W) onto the Removal Ring , loosen the Removal Ring by rotating it in the
direction of and pull the Bur Guard straight out.
_ Mounting
1
Align the Bur Guard and Attachment and tighten the
Removal Ring manually.
2
Rotate the Removal Ring in the direction of with the
Wrench for Slim/Super Slim (P200-1ARA-W) to fix it firmly.
8 Check before Operation
Conduct check before operation according to the following procedure to confirm that there are no abnormalities. If any abnormality is
observed, perform "11 Troubleshooting".
1
Attach the Attachment to the Motor, then attach the Cutting
Accessory to the attachment. Refer to Check before Operation
of each Attachment.
2
Check for abnormalities at the connection part of the
Attachment and Cutting Accessory.
3
Operate the Attachment at the rotational speed permissible for
the Cutting Accessory mounted. Confirm that nothing is loose
and that no abnormal vibration, heating or noise occurs.
If any abnormality is observed, stop use and perform "11
Troubleshooting".
4
Ensure that there are no abnormalities before use.

Table of Contents

Other manuals for NSK Primado2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NSK Primado2 and is the answer not in the manual?

NSK Primado2 Specifications

General IconGeneral
BrandNSK
ModelPrimado2
CategoryMedical Equipment
LanguageEnglish

Related product manuals